7T implants: Difference between revisions
From SpinozaWiki
No edit summary |
No edit summary |
||
Line 9: | Line 9: | ||
* implant anchored dentures (remove screw/click-on top part if possible) | * implant anchored dentures (remove screw/click-on top part if possible) | ||
* Nitinol coil (2019: experience UMCU) | * Nitinol coil (2019: experience UMCU) | ||
* stomach balloons (2022: Spatz 3 geen metaal en Obalon types klein titanium deel, getest op 3T. Prima voor brein scans. Ref UMCU mail) | |||
Line 17: | Line 18: | ||
* stents, coils, clips in cerebral vessels. | * stents, coils, clips in cerebral vessels. | ||
* permanent eyeliner (except for clincal patients) | * permanent eyeliner (except for clincal patients) | ||
* pacemaker leads of any type | |||
On a related note, the maximum weight for the bed of the 7T is 200 kilos (according to Anton/Philips manuals) | On a related note, the maximum weight for the bed of the 7T is 200 kilos (according to Anton/Philips manuals) |
Revision as of 11:33, 2 December 2022
This is a list of implants that have been approved for use at 7T by other labs. For now, they are ordered alphabetically.
- abdominal staples (2020: experience UMCU)
- amalgam fillings (2018: https://doi.org/10.1148/radiol.2018172597, discussed with UMCU toxicologist)
- dental wires/retainer (2013: study LUMC: https://onlinelibrary.wiley.com/doi/full/10.1002/mrm.25019)
- Essure IUD (2019: experience UMCU)
- hip prosthesis (for head scans)
- implant anchored dentures (remove screw/click-on top part if possible)
- Nitinol coil (2019: experience UMCU)
- stomach balloons (2022: Spatz 3 geen metaal en Obalon types klein titanium deel, getest op 3T. Prima voor brein scans. Ref UMCU mail)
Not approved following discussion with experts are:
- orthodontic braces (as opposed to the retainers above)
- spinal implants (https://www.spine-health.com/treatment/spinal-fusion/interbody-cages-spine-fusion)
- stents, coils, clips in cerebral vessels.
- permanent eyeliner (except for clincal patients)
- pacemaker leads of any type
On a related note, the maximum weight for the bed of the 7T is 200 kilos (according to Anton/Philips manuals)