3T hardware

  • Scanner: Philips Ingenia 3.0T CX
  • Bore diameter: 60 cm
  • Headcoil: ds Head 32channels 3.0T serial number 791 part number 453 5303 12384
  • Digital broadband architecture; digitizes the signal right in the coil.
  • Dual transmit parallel RF transmission technology
  • multiband SENSE (simultaneous multislice imaging) and compressed SENSE acceleration available
  • Max gradient strength: 40 mT/m
  • Max gradient slew rate: 200 mT/m/ms
  • Field-of-view (in head-feet direction): 45 cm

Setting up an MRI protocol

A basic MR scan session usually starts with the following scans:

  • Pre-scanning - a 3-plane localizer or 'scout' scan meant to find the subject's head. It is also be used for prescription for the subsequent scans. Doing some sort of localizer is necessary.
  • Anatomical MRI - usually a 3D T1-weighted scan at 1mm isotropic resolution. It is essential for image alignment and anatomical analysis. More choices of anatomical scans are listed in the Anatomical imaging section.
  • Shim - measures the magnetic field inhomogeneity and corrects it with polynomial gradients up to 2nd order. It should be run before fMRI or diffusion scans.

At this point you will want to add a number of functional scans, diffusion scans or other type of scans based on your experiment. In the next section we describe templates for different categories of MRI protocols. The protocol templates are organized by category.

You can get help in customizing the parameters from the Spinoza staff.

Pre-scanning

Anatomical MRI

T1 weighted

1.0mm, TR=2000ms, FOV 256x2256x176mm, flip angle=8°, TR=7ms, TE= 3ms, RCS = 6 (AP), oversampling (RL) 1.28, for a total of 38 shots, and a duration of 1:55sec

Please cite "Stutters et al., Structural 3DT1 scans with compressed sensing are suitable for cross-sectional brain volume measures in multiple sclerosis. Proc Intl Soc Magn Reson Med 29(2021): 2150"

T2 weighted

Classical BOLD fMRI

All with 10% gap, TE=27.63ms, SENSE 2, moderate PNS, maximum gradient mode, regular dynamic stabilization, 135Hz fat suppression SPIR offset, ascending slice order

3.0mm, TR=2000ms, FOV 240x240xxxxmm, WFS 12.0pix (36.0mm), Max dyns xxx (workhorse)
2.5mm, TR=3000ms, FOV 240x240xxxxmm, WFS 15.1pix (37.75mm), Max dyns xxx (high-res reduces signal drop out)

Multiband BOLD fMRI

All with 10% gap, TE=30ms, SENSE 1.5, multiband 4, moderate PNS, maximum gradient mode, regular dynamic stabilization, 175Hz fat suppression SPIR offset, ascending slice order

3.0mm, TR=0550ms, FOV 240x240x118mm, WFS 14.2pix (42.6mm), Max dyns 1820 (denoising in people with increased heart rate)
2.7mm, TR=0700ms, FOV 240x240x130mm, WFS 14.4pix (38.9mm), Max dyns 1489 (workhorse)
2.0mm, TR=1600ms, FOV=224x224x125mm, WFS 23.7pix (47.4mm), Max dyns 1170 (high-res reduces signal drop out) 

Considerations:

  • Set shimbox to contain as little non-brain tissue and air as possible or use image based shimming through MRCode tool (requires image_based_shim patch)
  • Create second GE-EPI with opposite fatshift direction, set preparation to auto for this sequence (so it won't perform a second B0 shim) and group the sequences
  • Perhaps not include dummies and use one of the pre-saturation scans for registration as it has better grey/white matter contrast (see Multiband data registration)

Diffusion Imaging

Magnetic Resonance Spectroscopy

If you have questions about our standard MRS sequences (e.g. GABA, glutamate), please contact Anouk Schrantee.