Quality Control procedure
The Quality Control (QC) procedure at the Spinoza centre is composed of a weekly and monthly protocol consisting various tests and checks. The QC is usually performed by the scan-assistants of the Spinoza centre, supervised by Tinka and Steven. All relevant data is saved into an excel-file. Below, both procedures are summarized.
Weekly quality control procedure
Scans:
- Philips Intelligent Quality Test (PIQT): quality check of spatial stability (uses 1 channel headcoil and Philips PQT phantom).
- BOLD-fMRI (EPI) and DTI-test using the AGAR-phantom: quality check of temporal stability (uses 32 channel headcoil)
- Spurious test: quality check of homogeneity of static magnetic field.
- QPI: check for spikes.
Check scanner computer:
- Virus scan (monthly) and updates (weekly)
Check stimulus-set up in scanner room:
- Set-up projection screen, beamer, and projected image
Check stimulus-computer:
- Volume settings, headphone sound quality, Presentation licence
- Trigger counter (during EPI-scan of the AGAR phantom) to check whether pulses are sent to the stimulus computer.
Check, cubicle computers (incl. NPO) and Mock scanner
- Mock scanner: check projection screen/image, headphones (and audio settings), response-boxes
Check, other:
- Back-up ProefpersonenRegistratie and Quality Control log (excel-sheets)
- Amount of screening-forms, information brochures, and "Toestemmingsverklaringen"
|
Monthly Quality Control procedure
Will be uploaded soon.