1: Respiratory sensor - 2: Finger clip - 3: Wireless Peripheral Pulse Unit (PPU) - 4: Wireless VCG-unit (VectorCardioGraphy) for ECG

1. Put a battery in both the PPU (3) and VCG (4) unit, even if you only use PPU. You may leave the VCG unit in the closet if you don’t use it.

Error creating thumbnail: File missing
Battery charger

2. Place the respiratory sensor (1) horizontally +/- 5 cm above the navel of your subject and secure it with the Velcro band.

3. Put the finger clip (2) on the left ring finger. Put the side with the light on the opposite side of the finger nail.

4. In the control room, press the right mouse button on the bottom left side of the screen when you are in the application software and choose PPU + Resp to show the measurements on the bottom left.

Error creating thumbnail: File missing
Press right mouse button and choose PPU + resp

5. The measurements are also shown on the little screen on the scanner.

Error creating thumbnail: File missing
Scanner with screen

To remove the battery from the unit, first push the battery handles on the sides before you pull out the battery.

Error creating thumbnail: File missing
Remove battery

The batteries last +/- 1 hour, so preferably change them for every participant and always put the used ones in the charger immediately.

Put all batteries back in the charger when you are finished scanning.

Your physiology data will be automatically exported with Flywheel. If you want to manually export your physiology data you can do this from G:/log/physlog or use the gtPacknGo software.