Important: Some features of the QA currently do not work, therefore please ignore the steps between Start Ignore and End Ignore
Preparation
- Make sure there is no patch activated
- Create a new patient using the following naming convention: QA_3T_yyyymmdd.
- Use the 3TQA scancard under SC > QA.
Start Ignore
- After loading the exam card, right click in or below the sequences to find the Edit Scan Control Parameters... option. Set the coil UI selection mode parameter (at the bottom) under the General tab to full manual. Click apply so that the setting is saved for the current login session (clicking Store will save this setting permanently!).
- Open the FBIRN sequence and find the coils tab. Make sure that Smart Select is turned off and that all elements are selected (blue) for both halves of the head coil.
End Ignore
Phantom placement
- Use the 32-channel head coil.
- Remove the cushioning inside the head coil to reveal two small depressions. Place the red base with the appropriate feet in the depressions and make sure that it's level.
- Place the FBIRN phantom on the red base. Make sure that the nozzle and sticker are facing straight up.
- Position the phantom so that the laser is at the same level as the nozzle.
Scanning
- Make sure all geolinks are set to TRA.
- Planning is the easiest on the FBIRN sequence due to its large window. Change the angulation by setting AP (deg), RL (deg) and FH (deg) to 0. Proceed to plan the scan so that the midlines are in the center of the phantom in all directions.
- Starting the scan will result in both PNS and SAR warnings, which may be ignored.
Post scanning
- Export the entire patient (in dicom format) to ftpsc_classic [er moet nog een project 3tqa aangemaakt worden op de ftp server, de exam card moet dan ook de string "3tqa" bevatten]
Start Ignore
- Open the Edit Scan Control Parameters... menu. Switch the coil UI selection mode parameter back to restricted.
End Ignore
- Helium check: Go to: Start>MR-Development>Display Helium level and press Measure Helium Level (Not during scanning!). Note down helium level in Notepad file under S: > scstaff > Forms and Procedures > Helium measurements
- Close the program, remove the phantom and log off