General safety aspects
- differences with 3T
Table operation
Er zijn diverse opties om te proberen het bed weer te activeren. Het beste is om die in deze volgorde te proberen:
- Stop/reset op handel uiteinde bed.
- Trek bed handmatig helemaal naar achteren. Dan 2 keer een 3 vingers reset op handel uiteinde bed. (duurt ca 2 minuten per keer)
- Undock bed manual door schakelaar op de grond naast de magneet omhoog te zetten. (undock) Daarna bed naar achteren trekken en op de rem zetten. Vervolgens schakelaar in onderste stand stand (dock) zetten. Let op want connector komt naar buiten in docking station!! Schakelaar in midden zetten (ready to dock) Connector gaat weer naar binnen. Tenslotte bed weer aansluiten. (duurt ook weer ca 2 minuten)
- CDAS reset (zie elders op deze wiki)
Als de tafel piept, is de bedpositie kwijt (ws veroorzaakt door resetten in verkeerde bedpositie). Druk op stop. Bed handmatig naar buiten schuiven. Reset. 2x 3-vinger reset.
!!!!!! Belangrijk!!!!! Bij al het resetten en aan/uit zetten, moet het tafelblad zover mogelijk naar achteren worden gezet.
Als er niets lukt is er nog de mogelijkheid om handmatig het blad naar binnen te duwen. Voorwaarde is dat de tafel wel gedocked is (connector aangesloten op het bed) en het blad los staat. Dat kan door op de stopknop te drukken. De rode lampen op de bekapping gaan aan en het blad kan dan vrij bewegen. Om hem naar binnen te schuiven moet je de beveiliging even omzetten door vooraan onder het bed Het handeltje even vast te houden tijdens het naar binnen duwen van het bed. Dit handmatig naar binnen duwen is uiteraard niet wenselijk en zwaar werk. Probeer niet al te snel te duwen, vooral als er een patient op ligt.
In alle gevallen graag Anton of mij bellen en als het mogelijk is willen we graag weten wat er gescand werd voordat het bed stopte. (protocol en tijd)
Useful commands at console
(might move to separate page accessible to all users)
- stop/start all processes (NB including data transfer!):
- permproc stop|start all stop|start all processes
- c:\nmr\gyroscan\commands\activate_configuration.pl. type y, Enter.
- start_stop_procs.pl stop also stops background processes. Needed in case of hardware reset in service room.
- Communication with PRIDE (e.g. for image-based shimming): PrideXS_XML. In case of error Access denied, no SW key pressed, kill process pridexsproc_win_cs.exe in Task Manager.
- Export raw data: use gt pack 'n go utility
- Location shim parameters: E:\Export\PRIDE\shimtool\linear_shim.txt
- Manually selecting coil elements:
- Control Parameter Editor - Scan (approx 1/3, 6x clicking):
- User def. channel selection - yes
- at initial, Xmit Coil selection. choose 0.
- Dual coil - No. (2e surface coil off, independent selection per surface coil not possible)
- Direct protocol comparison (useful while we still run on R3)
- Export exam card with the protocols you want to compare.
- Do this:
Delayed Reconstruction,
User def element selection: yes
selecteer elementen: waarde 1
Changed default settings
- Control parameter Reconstruction: Noise clip factor 1,2,3: 0.001. Avoids noise clipping of phase to -pi in case the SNR is too low.
- Increased maximum average power to 20: max_avg_pow: 20 in mpgcoil_mtx_tr.i, add file to patch folder to apply.
Power off and power on sequence(s)
CDAS
- Reset CDAS: turn off green switch. Wait 5 seconds.
- Turn red switch.
- Unplug sense-cable (on top, labeled with red arrow). Wait 5 seconds.
- Turn on in reverse order: red switch and green switch.
- Wait for bed-lights (top) to turn green. Now plug in the sense-cable (red arrow). This controls bore light.
- At bed console, set light to 50%.
Main System
- Power off sequence: 1. scan computer. 2. CDAS. 3 Power gradient (right) and System (left), see Leiden document (original in scstaff)
- Power on sequence: in reverse order
Compressors
Procedure testing/resetting compressors Helium pump: 2 small cubic devices in the far right corner. The green light on the left device should be flashing. Check messages on small LED-displays: scroll with arrow up/down buttons. Should all be OK, not ERR. Temperature of tubes should be below room temperature (i.e. cold). To reset: press OFF. You should hear it switching off. Wait, and press ON, again hearing it switching on. To hard reset: Press OFF. Wait. Turn small black switch from I to O. Wait. Turn back to I. Wait. Important: press ON to turn the compressor back on.
Common malfunctions
- table
- coldhead compressors
- gradient amplifiers
Philips contacts (?)
Cor Romijn Anton ? Merel ?