== Quality Control procedure ==
The Quality Control (QC) procedure at the Spinoza centre is composed of a weekly and monthly protocol consisting various tests and checks. The QC is usually performed by the scan-assistants of the Spinoza centre, supervised by Tinka and Steven. All relevant data is saved into an excel-file. Below, both procedures are summarized.
The Quality Control (QC) procedure at the Spinoza centre is composed of a weekly and monthly protocol consisting various tests and checks. The QC is usually performed by the scan-assistants of the Spinoza centre, supervised by Tinka and Steven. All relevant data is saved into an excel-file. Below, both procedures are summarized.


Line 6: Line 4:


'''Scans''':
'''Scans''':
{|
|


- Philips Intelligent Quality Test (PIQT): quality check of spatial stability (uses 1 channel headcoil).
* Philips Intelligent Quality Test (PIQT): quality check of spatial stability (uses 1 channel headcoil and Philips PQT phantom).
 
* BOLD-fMRI (EPI) and DTI-test using the AGAR-phantom: quality check of temporal stability (uses 32 channel headcoil)
- BOLD-fMRI (EPI) and DTI-test using the AGAR-phantom: quality check of temporal stability (uses 32 channel headcoil).
* Spurious test: quality check of homogeneity of static magnetic field.
 
* QPI: check for spikes.
- Spurious test: quality check of homogeneity of static magnetic field.
 
- QPI: check for spikes.




'''Check scanner computer''':
'''Check scanner computer''':


- Virus scan (monthly) and updates (weekly)
* Virus scan (monthly) and updates (weekly)




'''Check stimulus-set up in scanner room''':
'''Check stimulus-set up in scanner room''':


- Set-up projection screen, beamer, and projected image
* Set-up projection screen, beamer, and projected image


- Response boxes
* Response boxes


- GSR/Eyetracker batteries
* GSR/Eyetracker batteries




'''Check stimulus-computer''':
'''Check stimulus-computer''':


- Virus scan and update
* Virus scan and update


- Windows updates
* Windows updates


- Volume settings, headphone sound quality, Presentation licence
* Volume settings, headphone sound quality, Presentation licence


- Trigger counter (during EPI-scan of the AGAR phantom) to check whether pulses are sent to the stimulus computer.
* Trigger counter (during EPI-scan of the AGAR phantom) to check whether pulses are sent to the stimulus computer.




'''Check, cubicle computers (incl. NPO) and Mock scanner'''
'''Check, cubicle computers (incl. NPO) and Mock scanner'''


- Presentation license
* Presentation license


- Virus scan and update
* Virus scan and update


- Windows update
* Windows update


- Mock scanner: check projection screen/image, headphones (and audio settings), response-boxes
* Mock scanner: check projection screen/image, headphones (and audio settings), response-boxes




'''Check, other''':
'''Check, other''':


- Helium level
* Helium level


- Temperature
* Temperature


- Back-up ProefpersonenRegistratie and Quality Control log (excel-sheets)
* Back-up ProefpersonenRegistratie and Quality Control log (excel-sheets)


- Amount of screening-forms, information brochures, and "Toestemmingsverklaringen"
* Amount of screening-forms, information brochures, and "Toestemmingsverklaringen"
|}


== Monthly Quality Control procedure ==
== Monthly Quality Control procedure ==


Will be uploaded soon.
Will be uploaded soon.

Latest revision as of 15:14, 7 April 2015

The Quality Control (QC) procedure at the Spinoza centre is composed of a weekly and monthly protocol consisting various tests and checks. The QC is usually performed by the scan-assistants of the Spinoza centre, supervised by Tinka and Steven. All relevant data is saved into an excel-file. Below, both procedures are summarized.

Weekly quality control procedure

Scans:

  • Philips Intelligent Quality Test (PIQT): quality check of spatial stability (uses 1 channel headcoil and Philips PQT phantom).
  • BOLD-fMRI (EPI) and DTI-test using the AGAR-phantom: quality check of temporal stability (uses 32 channel headcoil)
  • Spurious test: quality check of homogeneity of static magnetic field.
  • QPI: check for spikes.


Check scanner computer:

  • Virus scan (monthly) and updates (weekly)


Check stimulus-set up in scanner room:

  • Set-up projection screen, beamer, and projected image
  • Response boxes
  • GSR/Eyetracker batteries


Check stimulus-computer:

  • Virus scan and update
  • Windows updates
  • Volume settings, headphone sound quality, Presentation licence
  • Trigger counter (during EPI-scan of the AGAR phantom) to check whether pulses are sent to the stimulus computer.


Check, cubicle computers (incl. NPO) and Mock scanner

  • Presentation license
  • Virus scan and update
  • Windows update
  • Mock scanner: check projection screen/image, headphones (and audio settings), response-boxes


Check, other:

  • Helium level
  • Temperature
  • Back-up ProefpersonenRegistratie and Quality Control log (excel-sheets)
  • Amount of screening-forms, information brochures, and "Toestemmingsverklaringen"

Monthly Quality Control procedure

Will be uploaded soon.